For full functionality of this site it is necessary to enable JavaScript.
EMIN.COM.LA
0
Product image

Sundy SDCHN636 ເຄື່ອງວິເຄາະຄາບອນໄຮໂດຣເຈນ ແລະ ໄນໂຕຣເຈນ (C(0.005%~100%), H(0.01%~50%),N(0.01%~50%))

ຕິດຕໍ່
ການຊຳລະເງິນທີ່ປອດໄພ
ການຮັບປະກັນຄຸນນະພາບ
ປ່ຽນ ແລະ ສົ່ງຄືນງ່າຍ
ມີບໍລິການຈັດສົ່ງ

Measuring Range: C(0.005%~100%) H(0.01%~50%) N(0.01%~50%)

Reproducibility: Cd≤1.00% Hd≤0.25% Nd≤0.15%

Repeatability: Cad≤0.5% Had≤0.15% Nad ≤ 0.08%

Auto loader: 36/72 Samples

Analysis time: ≤5min/Sample

Sample Mass: 75~105mg(Recommended:100mg)

Analysis Method: CH:Infrared Absorption N:Thermal Conduction

Combustion Supporting Gas: Oxygen Purity:≥99.5% Pressure:≥1MPa

Carrier Gas: Helium Purity:≥99.99% Pressure:≥1MPa

Driving Gas: Nitrogen / Dry Compressed Air Pressure:≥1MPa

Preheating Time: <2.5hour

Sample Adding/Replacing Function: Available

Auto Analysis: Yes

Crucible Replacing: Automatic

Furnace Temperature: Combination Furnace,950℃±5℃ Reduction Furnace:700℃±5℃

Air Tightness Detection: Automatic

Humidity: 20%to 85%, non-condensing

Ambient Temperature: 15 ℃ to 28℃

Power requirement: AC 220 V(-15%~10%)50/60Hz

Max.Power: 4.5kW

Dimensions: 710*650*1045mm

Weight: 140kg

Advance Calculation: Yes

Analysis Report: Yes

Statistics Report: Yes

System Monitor: Yes