Expected Lifetime (oxygen): 4-5 years
Oxygen Range: 0.0 to 100%
Oxygen Sensitivity: 0.0001
O2 Resolution: 0.0001
Analog Output: Not Available
Resolution: 0.0001
Accuracy: : +/- 1% of reading; +/- 0.1% absolute below 10.0%
Drift: less than 0.4% per year
Oxygen Sensor Type: Proprietary electrochemical cell, expected lifetime 4-5 years
Response Time: 15 seconds to T95
O2 Sensor Lifetime: 4-5 Years
O2 Sensor Body: ABS Resin
Sample Connection: Sample Probe
Sample Method: Water displacement method; alternately, product can be squeezed to introduce sample to analyzer
Sample Pump: Not Available
Pump Timing: Not Available
Minimum Sample Required: 0.5cc
Sampling Assembly: Probe with ultra fine zero volume needle, low volume PTFE tubing
Oxygen Calibration Controls: Fine SPAN, Coarse SPAN, Zero adjustment
Instrument Calibration: Calibrates with room air; coarse SPAN adjustment on back; fine span adjustment on side; ZERO adjustment on side
Exhaust port: Vents to air
Power Supply: 100-240V (50-60Hz) Charger included, with US or international interchangeable plug adapter (upon request)
Backup Power: Optional 12V internal battery pack
Battery Operation: Optional
Input Voltage: 100-240V, 50-60Hz
Battery Life: 8-12 hours before recharging
Measures: O2, CO2: CO2
Form Factor: Benchtop
Size: 9.8in x 4in x 10in. (249 x 102 x 254 mm)
Weight: 8 lb. (3.6 Kg)
Standards: CE, RoHS
Included Consumables: Low volume sample probe, three zero volume puncture needles, tubing & tee fitting, two plastic 1cc and 5cc syringes, 115/240Vac to 12Vdc power adapter
Included Documentation: Owners manual (includes operating instructions with detailed calibration instructions, calibration certficate, precautions), NIST calibration certificate
Origin of Goods: Our products are manufactured in the U.S.A.
Products by Gas: Oxygen
Products by Applications: Pharmacetical
Including
Fine .01% resolution standard
Includes sampling probe with ultra • low volume tubing
Includes fine needles
Easy to use side span adjustment
Optional Items
Optional internal battery. The battery pack is a 10 AA NiMH with high temperature safety shutoff, low voltage cutoff, and 16 hour typical runtime.
Optional internal pump (this allows analyzer to be used with pump actuation for applications with larger amounts of headspace)
CONTAINER TEST KIT – Add the container test kit for difficult applications where using the pump is not practical. Our patent pending sampling kit can be used to collect samples from almost any imaginable container. Simply puncture the package under water, collect the sample in our inverted funnel, and then sample via the rubber stopper.