Guaranteed Spec Range:

N₂O: 0.3 – 200 ppm

CH₄: 1.5 – 12 ppm

CO₂: 380 – 500 ppm

NH₃: 0 – 300 ppb

H₂O: 0 – 3 %

Ambient Temperature: 10 to 35°C

Ambient Humidity: <85% RH, non-condensing

Sample Pressure: 300 to 1,000 Torr (40 to 133 kPa)

Sample Flow Rate: ~230 sccm

Sample Humidity: <99% RH, non-condensing, water correction tested to 25°C dew point

Sample Temperature: -10 to 45°C

Cavity Temperature Control: ±0.005°C

Cavity Pressure Control: ±0.0002 atm

Closed-loop/Recirculation Capability: Compatible with Picarro Closed System Pump A0702

Inlet Fittings: 1/4” Swagelok®

Dimensions: 17” W × 7” H × 17.5” D (43.2 × 17.8 × 44.6 cm), not including 0.5” feet

Weight: 50 lbs (22.6 kg)

Power Requirements: 100–240 VAC; 47–63 Hz (auto-sensing); <375 W at start-up (total); steady-state operation: 120 W (analyzer), 150 W (pump)

Installation: Benchtop (standard) or 19” rack mount chassis (optional)

Accessories: Included: keyboard, mouse; Optional: LCD monitor; Excluded: vacuum pump

Options:

A0702: Picarro Closed System Pump

S0528: O₂ sensor for O₂ measurements and correction in varying O₂ environments

S0517: Extended CH₄ operating range up to 800 ppm