pH Tester Calibration Procedure

26/04/y 13:50:20

**1\. Range of application** This technical document specifies the calibration of pH tester with glass electrodes of (0-14) pH in measuring range and accurac

1. Range of application

This technical document specifies the calibration of pH tester with glass electrodes of (0-14) pH in measuring range and accuracy of at least 0,02 pH

2. Calibrating techniques

- Outside check

- Technical check

- Measurement check

3. Calibrating instruments

3.1 Standard measuring instrument (s)

Technical Calibration Buffer:


- Range of measurement: (0-14) pH

- Measurement Uncertainty: ≤ 0.01 pH

3.2 Other measuring instrument (s)

Heated Digital Water Bath:


- Range of measurement: (+5 ÷ 99) °C

Standard Humidity, Temperature Meter


- Range of measurement: (-20 ÷ 70) °C

- Accuracy:

  ± 0.2 °C (10 ÷ 40 °C)

  ± 0.4 °C of the remaining

  ± 1%RH (10 ÷ 90) %RH 

  ± 2.0 %RH (of the remaining)

Hygro-Thermometer

- Range of measurement: (0 ÷ 50) °C

- Accuracy: ±0.5 °C

- Range of measurement: (10 ÷95) %RH

- Accuracy: ±3%RH

3.3 Auxiliary Instrument(s)

- Distilled water

- Sprayer

- Tissue, glass cup

4. Conditions for Calibration

Temperature: (25 ± 2) °C

Humidity:  80 %RH

The temperature, as well as the humidity, must be recorded before and after calibration

5. Prepare for Calibration
Prior to the calibration, the following preparations must be done:
- Prepare the Technical Calibration Buffer
- Place the pH tester and Technical Calibration Buffer in the calibration room at least 1 hour prior to calibration. 
- The Technical Calibration Buffer must be kept at constant temperature at least 1 hr at (25 ± 0,2) 0C in Heated Digital Water Bath
- Prior to calibration, immerse the pH tester electrode with distilled water at least 3 hrs in the lab.
Before measuring the pH level of Technical Calibration Buffer, the followings must be met:
- Clean the pH tester’s electrodes with distilled water.
- Dry the electrode with tissue.
The pH tester to be calibrated must be operating at least 15 mins prior to calibration.

6. Begin the calibration
6.1 Outside Check
6.2 Technical Check
6.3 Measurement Check
A pH tester is checked and measured according to the following content, method and requirements:
- Calibrate the pH tester according to the manufacturer's technical documents.
- Check the error with standard pH solutions
- Measure at least 3 times for each pH standard solution. Record the measured results in the record.
- Calculate the error

7. Measurement Uncertainty Evaluation

8. General handling

a. The calibrated pH tester is stamped, certified in paper along with calibration results, and measurement uncertainty.

b. The recommended calibration cycle of pH tester is 12 months.

9. About EMIN's Calibration - Verification - Repair center

EMIN Vietnam’s Calibration and Verification center has been established since 2020 to provide after-sales services. The Calibration and Verification center is thoroughly equipped with technical equipment, followed by the assistance from various equipment manufacturers around the globe. At the center, calibration of specialized equipment is successfully carried out by experienced engineers in EMIN Vietnam, ensuring fast, accurate, reliable procedure to meet the most demanding customers’ requirements). Such technical capacity can only be performed by very few technology businesses in Vietnam. Meanwhile, after tireless effort, the Calibration and Verification center of EMIN Vietnam is officially certified to be accredited by the ISO 17025 standard (Jan, 2021)

If you want your instruments calibrated, please contact:

Equipment Verification - Calibration - Repair Service Center - EMIN Vietnam JSC 

Address: No 203 Nguyen Ngoc Vu Street, Trung Hoa Ward, Cau Giay District, Ha Noi City

Email: services@emin.vn - Hotline: 0968.060.220

Further information about pH tester calibration service can be found here.


ຂ່າວທີ່ກ່ຽວຂ້ອງ

What Is Resonant Frequency? Why Can a Single Frequency Range Damage the Entire Product?
10/07/y 15:10:46

A device may operate stably for months, yet suffer cracks, broken solder joints, or snapped component leads after being transported. What surprises many engineers is that the vibration amplitude during transport is no greater than usual. The root cause lies in a specific frequency range where the product's structure absorbs energy most intensely, triggering resonance and causing continuous stress concentration at a single point.

Should Carton Packaging Undergo Vibration Testing or Drop Testing First?
10/07/y 10:38:40

Một kiện hàng gần như không bao giờ bị rơi ngay khi vừa rời khỏi nhà máy. Quãng đường hàng trăm kilomet trên xe tải, container hay băng chuyền mới là giai đoạn diễn ra đầu tiên, nơi bao bì liên tục chịu rung lắc trong nhiều giờ. Chỉ sau đó mới đến những cú va chạm khi bốc xếp, giao nhận hoặc rơi ngoài ý muốn.

Distinguishing Between DFT and WFT in Coating Thickness Measurement
09/07/y 14:40:30

Many projects encounter a situation where a freshly applied coating looks perfectly uniform and paint consumption matches the estimate, yet additional coating is required during inspection because the thickness is insufficient. This leads many to suspect measurement errors or equipment inaccuracy, whereas the actual cause is a confusion between WFT (Wet Film Thickness) and DFT (Dry Film Thickness)

Can a car paint thickness gauge detect if a vehicle has been in a collision?
08/07/y 16:56:01

Many used car buyers often spend a considerable amount of time examining the paintwork, the gaps between body panels, or the overall condition of the exterior. However, these signs can be skillfully masked following repairs; precise color matching and careful polishing can make the vehicle appear almost factory-original.

ຕິດຕາມຂ່າວສານ ແລະ ຂໍ້ສະເໜີ

ຮັບສ່ວນຫຼຸດພິເສດຕາມປະລິມານ, ອັບເດດລາຄາຂາຍສົ່ງ ແລະ ການແຈ້ງເຕືອນສິນຄ້າໃໝ່ສົ່ງກົງເຖິງອິນບັອກຂອງທ່ານ.

ໂດຍການສະໝັກສະມາຊິກ, ທ່ານຍອມຮັບ ເງື່ອນໄຂການໃຫ້ບໍລິການ ແລະ ນະໂຍບາຍຄວາມເປັນສ່ວນຕົວ ຂອງພວກເຮົາ.

ການຊ່ວຍເຫຼືໍາດ່ວນ

ເຂົ້າເຖິງຜູ້ຊ່ຽວຊານທີ່ໄດ້ຮັບການຢັ້ງຢືນຂອງພວກເຮົາໂດຍກົງ