For full functionality of this site it is necessary to enable JavaScript.
EMIN.COM.LA
0
Product image

Mettler Toledo 52206395 DO Sensor InPro6970i/12/320

ຕິດຕໍ່
ການຊຳລະເງິນທີ່ປອດໄພ
ການຮັບປະກັນຄຸນນະພາບ
ປ່ຽນ ແລະ ສົ່ງຄືນງ່າຍ
ມີບໍລິການຈັດສົ່ງ

Accuracy: ± (1 % of the reading + 2 ppb)

Model Family: InPro 6900i optical series

CIP resistant: Yes

Response Time: T98 = < 20 s

Sensor Length: 320 mm

Cable Connector: M12 5-Pin

Design Pressure: 174 psi (12 bar)

Detection limit: 2 ppb

O-Ring Material: EPDM

Operating Range: 2 ppb – 2,000 ppb

Sensor Diameter: 12mm

Connector Design: Straight

Measurement Range: 0- 2000 ppb saturation

Process Connection: Pg 13.5

Digital Integration: RS485 - ISM®

Measurement Parameter: Oxygen

Measuring Temp. Range: 23 °F – 104 °F (-5 °C – 40 °C)

Measurement Technology: Optical fluorescence quenching

Mechanical Temp. Range: 23 °F – 250 °F (-5 °C – 121 °C)

ISM (Digital) or Analog: ISM (Digital)

Certificates & Approvals: Quality certificate, Material certificate 3.1, Surface finish certificate, FDA/USP Class IV

Operating Pressure (Bar): 0.2–12 Bar

Operating Pressure (PSI): 02.9–174 PSI

Wetted Metallic Material: 316L stainless steel

Wetted membrane material: Silicone

Autoclavable/Sterilizable: Yes

Primary Media/Application: Food and Beverage

Sterilizable/Autoclavable: Sterilizable

Measured Pressure Resistance: 0.2 bar – 12 bar

Residual signal in O2 free media: <0.025 of signal in ambient air

Hygienic/Biocompatibility Approvals: FDA/USP Class 7

Surface Roughnes of Wetted Material: N5/Ra16 (Ra=0.4μm/16μin)