Junray ZR-1015 Biosafety Cabinet Tester ຄຸນະພາບ (100L/min, 28000r/min)

ຕິດຕໍ່
ການຊຳລະເງິນທີ່ປອດໄພ
ການຮັບປະກັນຄຸນນະພາບ
ປ່ຽນ ແລະ ສົ່ງຄືນງ່າຍ
ມີບໍລິການຈັດສົ່ງ

Sampling flowrate: 100L/min  ±2.0%

- Resolution:  0.01L/min

Rotation speed of aerosol generator: 28000r/min ±50r/min

Highest position of X1,Y1: 1000mm

Disturbing cylinder: With horizontal measurement function, up to 1100mm

Noise: <65db(A)

Working temperature: ~40℃

Power supply: AC(220±22)V ,(50±1)Hz

Size: (length 321×width 240×height 175)mm

Weight(host): About 5.0kg

Power consumption:<400W

Standards

YY 0569-2011: Class II biosafety cabinet

NSF/ANSI 49-2020: Biosafety Cabinets Design and Performance

IEST-RP-CC007.3: Testing ULPA Filters

EN12469-2000: Biotechnology - Performance criteria for microbiological safety cabinets

JJF 1815-2020: Calibration specification for of class II biosafety cabinet


Features

• Good human-computer interaction

7.0 -inch color screen , touch operation.

Less accessories, simple installation and convenient operation.

One touch start-Four operation modes can be started with one touch.

Data recording function.Test results could be exported to USB drive or printed by printer.

• Automatic control, more accurate monitoring

Four path independent high precision sampling module, automatic flow control, don’t need to adjust pressure to stabilize flow.

The automatic aerosol control port adopts PID arithmetic, control by real time feedback generator rotation speed.