HV HIPOT GDC-9560B Gas Chromatograph Analyzer

ModelGDC-9560B
ຕົ້ນກຳເນີດChina
ການຮັບປະກັນ12 ເດືອນ
ຕິດຕໍ່
ການຊຳລະເງິນທີ່ປອດໄພ
ການຮັບປະກັນຄຸນນະພາບ
ປ່ຽນ ແລະ ສົ່ງຄືນງ່າຍ
ມີບໍລິການຈັດສົ່ງ

Temperature control range: room temperature 5-400℃, step 1℃, accuracy ±0.1℃

Temperature increasing steps: 16 steps.

Increasing speed: 0.1-40℃/min

Gas control: Mechanical valve control mode and electronic pressure flow control mode for option.

External event: 4 channels.

Sampler type: Packed column, capillary tube, six-way valve gas sampling, automatic head-space sampling for option.

Detector: FID, TCD, ECD(optional), FPD(optional)

Start mode: Automatic or manual.

Communication port: Ethernet, IEEE802.3


FID(flame ionization detector)

Detection limit: ≤ 2×10-11g/s (Hexadecane/isooctane)

Baseline noise: ≤ 5×10-14A

Baseline shift: ≤ 1×10-13A/30min

Linear range: ≥ 106

TCD(thermal conductivity detector)

Sensitivity: S ≥ 2500mV•ml/mg (benzene/methylbenzene) (enlarge 1, 2, 4, 8 times for option)

Baseline noise: ≤ 20μV

Baseline shift: ≤ 30μV/30min

Linear range: ≥ 104


ECD (electron capture detector) optional

Detection limit: 1×10-13g/ml (Hexachlorocyclohexane/isooctane)

Baseline noise: ≤ 0.03mV

Baseline shift: ≤ 0.2mV/30min

Linear range: 103

Radioactive source: 63Ni

FPD (flame photometric detector) optional

Detection limit: (S) ≤5×10-11g/s, (P)≤1×10-12g/s(parathion-methyl/ absolute ethyl alcohol)

Baseline noise: ≤ 3×10-13A

Baseline shift: ≤ 2×10-12A/30min

Linear range: Sulphur ≥ 102, phosphorus ≥ 103

Datasheet