For full functionality of this site it is necessary to enable JavaScript.
EMIN.COM.LA
0
Product image

Gastec 139 Quick-measuring Detector tubes 1,2-Dichloroethylene ClCH:CHCl (5~250ppm)

ຕິດຕໍ່
ການຊຳລະເງິນທີ່ປອດໄພ
ການຮັບປະກັນຄຸນນະພາບ
ປ່ຽນ ແລະ ສົ່ງຄືນງ່າຍ
ມີບໍລິການຈັດສົ່ງ

Indicator: 10 to 100 ppm
Measuring Range:
+ 5 to 10 ppm, Number of Pump Strokes 2
+ 10 to 100 ppm, Number of Pump Strokes 1
+ 100 to 250 ppm, Number of Pump Strokes 1/2
Sampling Time: 45 seconds per pump stroke
Colour Change: Yellow → Reddish Purple
Reaction Principle: ClCH:CHCl + PbO2 + H2SO4 → HCl, HCl + Base → Chloride
Coefficient of Variation: 10% (10 ~ 20 ppm), 5% (20 ~100 ppm)
Nr. of tests: 10
Pump: GV-100S/110S

Catalog

Manual

Performance
Measuring Range
5 to 10 ppm 10 to 100 ppm 100 to 250 ppm
Number of Pump Strokes
2 1 1/2
Correction Factor
1/2 1 2.5
Sampling Time 45 seconds per pump stroke
Detecting Limit 1 ppm (n=2)
Colour Change Yellow → Reddish Purple
Reaction Principle

ClCH:CHCl + PbO2 + H2SO4 → HCl
HCl + Base → Chloride

Coefficient of Variation 10% (for 10 to 20 ppm), 5% (for 20 to 100 ppm)
Shelf Life 2 Years
Corrections for temperature & humidity Temperature correction is necessary
Store the tubes in the refrigerator to keep at 10ºC (50ºF) or below.
Possible coexisting substances and their interferences (NOTE)
Substance Concentration Interference Change colour by itself to
Nitrogen oxides   No No
Hydrogen chloride, Halogens   + Reddish purple
Tetrachloroethylene   + Reddish purple
Trichloroethylene   + Reddish purple
Calibration gas generation Diffusion tube method
TLV-TWA TLV-STEL Explosive range
200 ppm - -
ຕິດຕາມຂ່າວສານ ແລະ ຂໍ້ສະເໜີ

ຮັບສ່ວນຫຼຸດພິເສດຕາມປະລິມານ, ອັບເດດລາຄາຂາຍສົ່ງ ແລະ ການແຈ້ງເຕືອນສິນຄ້າໃໝ່ສົ່ງກົງເຖິງອິນບັອກຂອງທ່ານ.