Range: –80 °C to 110 °C
Stability:
±0.006°C at –80°C (ethanol)
±0.005°C at 0°C (ethanol)
±0.005°C at 100°C (oil 5012)
Uniformity:
±0.007 °C at –80 °C (ethanol)
±0.007 °C at 0 °C (ethanol)
±0.007 °C at 100 °C (oil 5012)
Heating Time: 60 minutes, from 25 °C to 100 °C (oil 5012)
Cooling Time: 210 minutes, from 25 °C to –80 °C (ethanol)
Stabilization Time: 15–20 minutes
Temperature Setting: Digital display with push-button data entry
Set-Point Resolution: 0.01°; 0.00018° in high-resolution mode
Display Resolution: 0.01 °
Digital Setting Accuracy: ±1°C
Digital Setting Repeatability: ± 0.01 °C
Access Opening: 120 x 172 mm (4.7 x 6.8 in)
Depth:
457 mm (18 in) without Liquid-in-Glass Thermometer Cal Kit
482 mm (19 in) with Liquid-in-Glass Thermometer Cal Kit
Wetted Parts: 304 stainless steel
Power: 230 VAC (±10 %), 50 or 60 Hz, specify, 10 A
Volume: 15.9 liters
Size (WxDxH): 1067 x 356 x 788 mm (940 mm from floor to tank access opening)
Weight: 91 kg (200 lb.)