X-ray tube voltage Vp range : Single energy, B models: 40–160kVp; Single energy, C models: 80–225kVp; Dual energy, D models: 120–225kVp
Scintillator material : Single energy, B models: GOS screen; Single energy, C models: Ceramic; Dual energy, D models: Fast GOS & Ceramic
Active area lengths, single module : 672, 1152, 1344, 1442 mm
Total pixel number per module : Single energy: 448, 768, 896. Dual energy: 896,1536,1792
Maximum number of modules and active length with one XGCU EX-PH control unit: 5 modules, 6720 mm
Pixel size (photodiode) : 1.4 x 3.2 mm
Scanning speed:
Maximum scanning speed : Single energy: 5 m/s. Dual energy: 2.5 m/s
Integration time range : 0.58…128 ms
Line averaging and summing : Yes, up to 256x
Pixel binning : Yes. 1, 2 or 4 pixels
A/D resolution : 16 bits
Overall X-ray response non-uniformity, dark offset subtracted at detector level : < ±25 %
X-ray response linearity : >99 %
Max number of dead pixels : Zero, no dead pixels
Channel electronic crosstalk: < 0.5 %
Dynamic range : >16 000:1 depending on sensitivity settings
Sensitivity settings : Range 7.5..84 pC with 7.5 pC steps (12 settings)
Interface, detector module to control unit : X-Link protocol in Cat5e cable with RJ45 connectors
Interface, control unit to host computer : Gigabit Ethernet, UDP protocol
EMC compliance : EN61326-1, EN61000-3-2, EN61000-3-3
IP classification : IP55
RoHS compliance: Yes
Operational voltage and power : 24 VDC, max 35 W per module. Industrial power supply units available from DT
Operational temperature and humidity : 0–40 °C, 30–80 % (non-condensing)
Storage temperature : -10–50 °C
Lifetime under X-ray : typical 100 kGy
Weight per module : 5 to 40 kg
Power on time recording : Yes, up to 100 000 hours