INPUT SPECIFICATIONS:
Voltage: 100 – 120 VAC / 200 – 240 VAC ± 10% Auto Range
Frequency: 50/60 Hz ± 5%
Fuse: 3.15 A, Fast Blow 250 VAC
DIELECTRIC WITHSTAND TEST MODE
Output Rating:
5 kVA @ 20 mAAC
6 kVA @ 7.5 mADC
Maximum Limit:
AC
Range: 0.00 – 20.00 mA
Resolution: 0.01 mA
DC
Range: 0 – 7500 µA
Resolution: 1 µA
Accuracy: AC and DC ± (2% of setting + 2 counts
Minimum Limit:
AC
Range: 0.000 – 9.999 mA
Resolution: 0.001 mA
DC
Range: 0.0 – 999.9 µA
Resolution: 0.1µA
Accuracy: AC and DC ± (2% of setting + 2 counts)
Arc Detection:
Range: 1-9, ON/OFF Select
Ground Fault Interrupt
GFI Trip Current: 450 µA max (AC or DC), Fixed HV Shut Down Speed: < 1 msec
Current Display:
AC
Range 1: 0.000 – 4.000 mA
Range 2: 3.50 – 20.00 mA
DC
Range 1: 0.0 µA – 400.0 µA
Range 2: 0.350 mA – 4.000 mA
Range 3: 3.50 mA – 7.50 mA
Accuracy: All Ranges ± (2% of reading + 2 counts)
DC Output Ripple: ≤ 5% Ripple rms at 6 kVDC @ 7.5 mA Resistive Load
RAMP-HI Selectable:
Range: 0.0 – 7,500 µA, User Selectable
Charge-LO: 0 – 350 µA DC or Auto Set
Ground Continuity
Range: 0.00 – 1.50 Ω
Resolution: 0.01 Ω
Accuracy: ± (3% of setting + 0.02 Ω)
Ground Continuity Auto Offset:
Range: 0.00 – 0.50 Ω
Resolution: 0.01 Ω
Accuracy: ± (3% of setting + 0.02 Ω)
Voltage Setting
Range: 30 – 1,000 VDC
Resolution: 1 V
Accuracy: ± (2% of setting + 5 V)
Resistance Display: Range: 1 – 50,000 MΩ
Remote Control and Signal I/O:
Inputs: Test, Reset, Hardware Interlock, File Recall
Outputs: Pass, Fail, Test-in-Process, Reset-Out, Start-Out
Vmax: Displays the maximum voltage value recorded during a breakdown
Imax: Displays the maximum leakage current value read during a test
Memories:
50 steps
1500 test results
Interface: USB standard
Language: English, Traditional Chinese, Simplified Chinese, Turkish, Portuguese, Spanish, German, French
Security: Multiple user setups with ID and password